> 40 CTC Analytics Consumables

X-Type Syringes from CTC Analytics

MicroLiter continues to support all aspects of the CTC Analytics PAL®

Sample Handler. We are pleased to assist all customers with trouble-

shooting consumable issues and we encourage you to call us regarding

any inquiry you may have. While we believe that there are some

excellent alternatives to the CTC endorsed syringes, we have moved

away from offering all product lines for syringes available for the PAL®

and will focus our efforts on only the Hamilton products developed by

CTC Analytics.

MicroLiter has long been a partner for consumables in the USA with

LEAP Technologies and will continue to seek opportunities to grow with

this dynamic innovator for sample handling solutions.

Our move to reduce our syringe offerings in no way reflects an overall

change in our relationship with LEAP or how together we assist our

mutual customers in finding solutions for their particular requirements.

General Specifications

Within these listings are fitted and gastight plungers, fixed needles with

various point and gauge configurations. Use the chart below to choose

the point style required for your application. Please refer to the specific

instrument listings on the following pages for easier selection.

Point Styles:

Gauge:

3 Blunt (straight) cut for LC

22 Large Outer Diameter

5 Side hole needle, with a sharp

22s Large Outer Diameter

conical solid tip for GC

Low Dead-Volume

AS Conical Tip with 8° bevel for GC 23 Large Outer Diameter

for Merlin MicroSeal

26 Small Outer Diameter

26s Small Outer Diameter

Low Dead-Volume

Plunger Styles:

Application:

Fitted Stainless Steel Plunger

GC and Limited LC Applications

PTFE Tipped

GC and Volatile LC

Polymer Tipped

Low Volatility High-Throughput LC

LC Samplers

For LC work, a needle gauge of 22 with a blunt tip (point style 3) is

required to match the valve inlet. 22s gauge needle specifications are

recommended to reduce coring of septa when possible.

GC Samplers

For GC work we recommend a small diameter, low dead-volume needle

(26s = 26 gauge with a low dead volume). The result will be: easy

purging of air, high reproducibility and low cleaning solvent use.

Needle point styles must be considered when piercing a GC septum.

One-side PTFE and one-side Rubber or Silicone septa commonly used

for GC applications require a beveled conical point style AS or side-hole

needle (point style 5) to prevent coring.

Gastight plungers are only essential for very sticky samples and / or

highly volatile solvents like hexane. They are also needed at syringe

volumes above 25µl.

Merlin MicroSeal™

requires a 23 gauge needle with point style AS so

that a proper seal is achieved during injection in the GC inlet.

MicroLiter