Previous Page  111 / 196 Next Page
Information
Show Menu
Previous Page 111 / 196 Next Page
Page Background

WHEATON

800.225.1437 (U.S. & Canada Only) | 856.825.1100 | 856.825.1368 (F) |

www.wheaton.com

> 110

Liquid Handling

Cat. No.

Cat. No.

Inaccuracy (E%)

Imprecision (CV%)

Initial Package Pipette Alone

Volume

Division

Min. vol.

Mid. Vol.

Max. Vol.

Min. Vol.

Mid. Vol.

Max. Vol.

Tip Style

Micropipettes Acura

®

electro

926XS

W870902-A W870902

0.1 - 2µL

0.01µL

<± 2.5 %

1)

<± 1.2 % <± 0.9 % < 2.5 %

1)

< 1.5 % < 0.8 %

10µL

W870904-A W870904

0.5 - 10µL

0.05µL

<±1.2 %

2)

<± 0.8 % <± 0.6 % < 1.5 %

2)

< 0.7 % < 0.35 %

10µL

W870906-A W870906

0.5 - 10YµL*

0.05µL

<± 1.2 %

2)

<± 0.8 % <± 0.6 % < 1.7 %

2)

< 0.8 % < 0.4 %

200µL

W870908-A W870908

1 - 20µL

0.1µL

<± 1.2 %

2)

<± 0.6 % <± 0.5 % < 1.2 %

2)

< 0.4 % < 0.3 %

200µL

W870910-A W870910

2.5 - 50µL

0.25µL

<± 1.0 %

2)

<± 0.6 % <± 0.5 % < 0.7 %

2)

< 0.3 % < 0.25 %

200µL

W870912-A W870912

5 - 100µL

0.5µL

<± 1.0 %

2)

<± 0.6 % <± 0.5 % < 0.7 %

2)

< 0.3 % < 0.2 %

200µL

W870914-A W870914

10 - 200µL

1.0µL

<± 1.0 %

2)

<± 0.6 % <± 0.4 % < 0.6 %

2)

< 0.2 % < 0.15 %

200µL

W870916-A W870916

50 - 1000µL

5.0µL

<± 0.8 %

2)

<± 0.5 % <± 0.4 % < 0.4 %

2)

< 0.15 % < 0.1 %

1000µL

Macropipettes Acura

®

electro

936

W870918-A W870918

0.1 - 2mL

0.01mL

<± 1.5 %

2)

<± 1.0 % <± 0.5 % < 0.6 %

2)

< 0.3 % < 0.15 %

2mL

W870920-A W870920

0.25 - 5mL

0.05mL

<± 1.2 %

2)

<± 0.8 % <± 0.5 % < 0.6 %

2)

< 0.3 % < 0.15 %

5mL

W870922-A W870922

0.5 - 10mL

0.05mL

<± 1.0 %

2)

<± 0.7 % <± 0.5 % < 0.5 %

2)

< 0.2 % < 0.15 %

10mL

8-Channel Pipettes Acura

®

electro

956

W870924-A W870924

0.5 - 10µL

0.05µL

<± 3.5 %

2)

<± 1.5 % <± 1.0 % < 3.0 %

2)

< 0.9 % < 0.7 %

10µL

W870926-A W870926

2.5 - 50µL

0.25µL

<± 1.0 %

2)

<± 0.9 % <± 0.8 % < 1.0 %

2)

< 0.6 % < 0.4 %

200µL

W870928-A W870928

10 - 200µL

1.0µL

<± 0.9 %

2)

<± 0.7 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

200µL

W870930-A W870930

20 - 350µL

5.0µL

<± 1.0 %

2)

<± 0.8 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

350µL

12-Channel Pipettes Acura

®

electro

956

W870932-A W870932

0.5 - 10µL

0.05µL

<± 3.5 %

2)

<± 1.5 % <± 1.0 % < 3.0 %

2)

< 0.9 % < 0.7 %

10µL

W870934-A W870934

2.5 - 50µL

0.25µL

<± 1.0 %

2)

<± 0.9 % <± 0.8 % < 1.0 %

2)

< 0.6 % < 0.4 %

200µL

W870936-A W870936

10 - 200µL

1.0µL

<± 0.9 %

2)

<± 0.7 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

200µL

W870938-A W870938

20 - 350µL

5.0µL

<± 1.0 %

2)

<± 0.8 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

350µL

Performance values obtained in Forward mode with double-distilled water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655.

Performance measured at:

1)

0.5 µL,

2)

10% of nominal volume

* This pipette must be used with a 200µL pipette tip

Ordering Information

Initial package includes pipette, charging stand and accessories. Additional pipettes (alone) supplied with individual QC certificate, pipette tips samples and operating instructions.

Micropipettes

Reduced shaft length and conical end improve driveability

Easy access to microtubes and microplates

Optional longer volumetric modules

Simple conversion by user at any time

Unsurpassed metrological performance and stability

Macropipettes

Interchangeable nozzle protection filter

Adapters for Pasteur pipette and straw tips

Outstanding accuracy and precision

Multichannel

Large selection of 8 and 12-channel models

Lightweight and perfect hand fitting

Sequential tip ejection

Also fits any single channel volumetric module

Acura

®

electro

926 XS / 936 / 956 Electrontic Pipettes