Previous Page  111 / 196 Next Page
Information
Show Menu
Previous Page 111 / 196 Next Page
Page Background

WHEATON

800.225.1437 (U.S. & Canada Only) | 856.825.1100 | 856.825.1368 (F) |

www.wheaton.com

> 110

Liquid Handling

Cat. No.

Cat. No.

Inaccuracy (E%)

Imprecision (CV%)

Initial Package Pipette Alone

Volume

Division

Min. vol.

Mid. Vol.

Max. Vol.

Min. Vol.

Mid. Vol.

Max. Vol.

Tip Style

Micropipettes Acura

®

electro

926XS

W870902-A W870902

0.1 - 2µL

0.01µL

<± 2.5 %

1)

<± 1.2 % <± 0.9 % < 2.5 %

1)

< 1.5 % < 0.8 %

10µL

W870904-A W870904

0.5 - 10µL

0.05µL

<±1.2 %

2)

<± 0.8 % <± 0.6 % < 1.5 %

2)

< 0.7 % < 0.35 %

10µL

W870906-A W870906

0.5 - 10YµL*

0.05µL

<± 1.2 %

2)

<± 0.8 % <± 0.6 % < 1.7 %

2)

< 0.8 % < 0.4 %

200µL

W870908-A W870908

1 - 20µL

0.1µL

<± 1.2 %

2)

<± 0.6 % <± 0.5 % < 1.2 %

2)

< 0.4 % < 0.3 %

200µL

W870910-A W870910

2.5 - 50µL

0.25µL

<± 1.0 %

2)

<± 0.6 % <± 0.5 % < 0.7 %

2)

< 0.3 % < 0.25 %

200µL

W870912-A W870912

5 - 100µL

0.5µL

<± 1.0 %

2)

<± 0.6 % <± 0.5 % < 0.7 %

2)

< 0.3 % < 0.2 %

200µL

W870914-A W870914

10 - 200µL

1.0µL

<± 1.0 %

2)

<± 0.6 % <± 0.4 % < 0.6 %

2)

< 0.2 % < 0.15 %

200µL

W870916-A W870916

50 - 1000µL

5.0µL

<± 0.8 %

2)

<± 0.5 % <± 0.4 % < 0.4 %

2)

< 0.15 % < 0.1 %

1000µL

Macropipettes Acura

®

electro

936

W870918-A W870918

0.1 - 2mL

0.01mL

<± 1.5 %

2)

<± 1.0 % <± 0.5 % < 0.6 %

2)

< 0.3 % < 0.15 %

2mL

W870920-A W870920

0.25 - 5mL

0.05mL

<± 1.2 %

2)

<± 0.8 % <± 0.5 % < 0.6 %

2)

< 0.3 % < 0.15 %

5mL

W870922-A W870922

0.5 - 10mL

0.05mL

<± 1.0 %

2)

<± 0.7 % <± 0.5 % < 0.5 %

2)

< 0.2 % < 0.15 %

10mL

8-Channel Pipettes Acura

®

electro

956

W870924-A W870924

0.5 - 10µL

0.05µL

<± 3.5 %

2)

<± 1.5 % <± 1.0 % < 3.0 %

2)

< 0.9 % < 0.7 %

10µL

W870926-A W870926

2.5 - 50µL

0.25µL

<± 1.0 %

2)

<± 0.9 % <± 0.8 % < 1.0 %

2)

< 0.6 % < 0.4 %

200µL

W870928-A W870928

10 - 200µL

1.0µL

<± 0.9 %

2)

<± 0.7 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

200µL

W870930-A W870930

20 - 350µL

5.0µL

<± 1.0 %

2)

<± 0.8 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

350µL

12-Channel Pipettes Acura

®

electro

956

W870932-A W870932

0.5 - 10µL

0.05µL

<± 3.5 %

2)

<± 1.5 % <± 1.0 % < 3.0 %

2)

< 0.9 % < 0.7 %

10µL

W870934-A W870934

2.5 - 50µL

0.25µL

<± 1.0 %

2)

<± 0.9 % <± 0.8 % < 1.0 %

2)

< 0.6 % < 0.4 %

200µL

W870936-A W870936

10 - 200µL

1.0µL

<± 0.9 %

2)

<± 0.7 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

200µL

W870938-A W870938

20 - 350µL

5.0µL

<± 1.0 %

2)

<± 0.8 % <± 0.6 % < 0.6 %

2)

< 0.4 % < 0.25 %

350µL

Performance values obtained in Forward mode with double-distilled water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655.

Performance measured at:

1)

0.5 µL,

2)

10% of nominal volume

* This pipette must be used with a 200µL pipette tip

Ordering Information

Initial package includes pipette, charging stand and accessories. Additional pipettes (alone) supplied with individual QC certificate, pipette tips samples and operating instructions.

Micropipettes

■

Reduced shaft length and conical end improve driveability

■

Easy access to microtubes and microplates

■

Optional longer volumetric modules

■

Simple conversion by user at any time

■

Unsurpassed metrological performance and stability

Macropipettes

■

Interchangeable nozzle protection filter

■

Adapters for Pasteur pipette and straw tips

■

Outstanding accuracy and precision

Multichannel

■

Large selection of 8 and 12-channel models

■

Lightweight and perfect hand fitting

■

Sequential tip ejection

■

Also fits any single channel volumetric module

Acura

®

electro

926 XS / 936 / 956 Electrontic Pipettes