WHEATON
800.225.1437 (U.S. & Canada Only) | 856.825.1100 | 856.825.1368 (F) |
www.wheaton.com> 117
Liquid Handling
Acura
®
manual
815 / 835 Fixed Volume Pipettes
Fixed volume micro and macropipettes show the most stable performance. They provide
consistent results for any analytical or routine diagnostic tests by eliminating the risk of
erroneous volume setting.
Product Advantages
■
Slim ergonomic design and lightweight
■
Soft plunger activation
■
Justip
™
adjustable tip ejector* fitting most tips
■
User calibration with integrated key
■
Color coded smartie cap
■
Selection includes macropipette volumes
■
Three-year warranty
* Socorex patented
815 / 835F
1µL to 10mL
Tip Ejection - Easier Than Ever
Most ergonomically positioned, the large surface, soft padded ejector button provides for
low-pressure activation. In addition to ease of tip ejection, the shaft height adjustment
system called Justip
™
- controlled by efficient click-stops - allows a wide selection of
tips to tightly fit the nozzle.
Cat. No.
Volume
Inaccuracy (E%)
Imprecision (CV%)
Tip Style
Acura
®
manual
815
W810000
1µL
< ± 2.0 %
< 2.0 %
10µL
W810002
5µL
< ± 1.4 %
< 1.0 %
10µL
W810004
10µL
< ± 0.7 %
< 0.7 %
10µL
851206
10YµL*
< ± 0.7 %
< 0.8 %
200µL
W810006
15µL
< ± 0.7 %
< 0.5 %
200µL
W810008
20µL
< ± 0.75 %
< 0.5 %
200µL
W810010
25µL
< ± 0.7 %
< 0.6 %
200µL
W810012
30µL
< ± 0.7 %
< 0.6 %
200µL
W810013
32µL
< ± 0.7 %
< 0.6 %
200µL
W810014
40µL
< ± 0.7 %
< 0.5 %
200µL
W810016
50µL
< ± 0.7 %
< 0.5 %
200µL
W810018
60µL
< ± 0.7 %
< 0.6 %
200µL
W810020
70µL
< ± 0.7 %
< 0.6 %
200µL
W810022
75µL
< ± 0.7 %
< 0.5 %
200µL
W810024
80µL
< ± 0.7 %
< 0.4 %
200µL
W810026
90µL
< ± 0.7 %
< 0.3 %
200µL
W810028
100µL
< ± 0.7 %
< 0.3 %
200µL
W810030
120µL
< ± 0.7 %
< 0.4 %
200µL
W810032
150µL
< ± 0.7 %
< 0.3 %
200µL
W810034
200µL
< ± 0.7 %
< 0.4 %
200µL
W810036
250µL
< ± 0.7 %
< 0.4 %
1000µL
W810038
300µL
< ± 0.7 %
< 0.4 %
1000µL
W810040
400µL
< ± 0.7 %
< 0.3 %
1000µL
W810042
500µL
< ± 0.7 %
< 0.3 %
1000µL
W810044
600µL
< ± 0.7 %
< 0.3 %
1000µL
W810046
700µL
< ± 0.7 %
< 0.2 %
1000µL
W810048
750µL
< ± 0.7 %
< 0.2 %
1000µL
W810050
800µL
< ± 0.7 %
< 0.2 %
1000µL
W810052
900µL
< ± 0.7 %
< 0.2 %
1000µL
W810054
1000µL
< ± 0.6 %
< 0.2 %
1000µL
Acura
®
manual
835F
W810056
2mL
< ± 0.8 %
< 0.5 %
2mL
W810058
2.5mL
< ± 0.8 %
< 0.5 %
5mL
W810060
5mL
< ± 0.7 %
< 0.3 %
5mL
W810062
10mL
< ± 0.5 %
< 0.2 %
10mL
Performance values obtained with double-distilled water at constant temperature (± 0.5°C)
comprised between 20 and 25°C in accordance with ISO 8655.
* This pipette must be used with a 200µL pipette tip
Performance and Ordering Information