

WHEATON
800.225.1437 (U.S. & Canada Only) | 856.825.1100 | 856.825.1368 (F) |
www.wheaton.comLiquid Handling
> 131
How To Use a Pipette
Abstract
Pipettors are precision tools that, when used properly, will provide years of trouble-free
service. However, many users do not receive training to use this common laboratory
utensil.
Procedure
Socorex
®
Pipettors are precision tools that will help you efficiently perform your research
and testing tasks. Like any tools, your pipettor should be used and stored with care to
provide the best possible results.
The information below applies to Socorex Pipettors as well as other manufacturers'
pipettors. All air displacement pipettors work on the same principle. A plunger is
used to displace air- when the plunger is returned to its initial position (upper stop),
it creates a vacuum that is used to draw liquid into the disposable tip. The pipette
is calibrated such that the vacuum draws a volume of water as indicated on the
volume setting. Liquids lighter than water are pulled further into the tip, so these liquids
will pipette to a larger volume. Similarly, liquids heavier than water pipette to a lesser
volume. In the case of either lighter or heavier liquids, compared with water, the volume
dispensed will be different than that indicated on the pipettor. Pushing the plunger
pushes air, which then pushes on the liquid in the tip, which is dispensed.
It is important that you use approved tips with your pipettors to assure accuracy. Using
unapproved tips may lead to inconsistent results.
Normal Pipetting
1. Fit the tip, set the volume (variable pipettors only) and press down the plunger to the
first stop (metering stroke) with your thumb. Immerse the tip 2-3mm in the sample
while holding the pipettor vertically.
2. Slowly retract the pipetting plunger while watching the liquid fill the tip. You should
not observe bubbles or turbulence, which indicate gasses being pulled from the
liquid. These gasses affect the vacuum that draws the liquid, reducing the amount of
liquid aspirated into the pipet tip.
3. When the pipetting plunger has been retracted to its upper stop, remove your thumb
as the absence of pressure increases the precision of the pipettor. Slowly withdraw
the pipettor from the liquid. Wipe any drops on the outside of the tip on the wall of the
vessel you are drawing liquid from.
4. To dispense the liquid, hold the tip against the side of the receiving container at a
slight angle. Use your thumb to push down the pipetting plunger to the first stop, and
hold it for one second. After one second, push the button to the second stop. Pushing
to the second stop blows out any liquid left in the tip.
For work with volatile solvents, such as methanol, you may perform this procedure on
a “dummy” sample to saturate the system with vapor. This may improve accuracy, as
pre-saturating the air will reduce the tendency to “blow out” the liquid before you are
ready to dispense the sample. You should saturate the tip in this fashion every time you
change tips.
Reverse Pipetting
Reverse pipetting is used to aspirate an additional volume of liquid. This technique is
useful when working with thick, viscous liquids. This is also useful for volatile solvents.
1. Press the pipetting plunger with your thumb to the second stop. This is different from
the procedure listed for normal pipetting.
2. Holding the pipettor vertically, slowly retract the plunger to its upper stop. Wait for
the liquid to properly fill the tip. With viscous liquids, this will take longer than when
pipetting water. A larger amount of liquid will be aspirated than normal operation
since the plunger was pushed to the second stop.
3. When dispensing, push the plunger only as far as the first stop. Wipe any liquid
hanging on the tip on the side of the receiving container. Any remaining liquid will be
discarded with the tip.
Working Position
When aspirating the sample, the pipettor must be held vertically, or else too much liquid
will be drawn in. Tilting the pipettor by 30 degrees causes nearly 1% more liquid to be
drawn!
When dispensing the sample, the tip should be held at an angle against the container to
draw out the liquid in the tip. Under normal pipetting operations, analytical chemists will
recognize the pipettor as a “to contain” pipettor.
When aspirating the sample, the tip should generally be immersed to 2-3mm. Placing
the tip deeper into the sample allows pressure from the liquid to help push the sample
into the tip, reducing accuracy.
Working Conditions
Under ideal conditions, the sample should have the same temperature used to calibrate
the pipettor. Cold liquids are denser than warm liquids. If the pipettor was calibrated at
room temperature, but used in a cold-room, smaller samples than expected would be
dispensed.
Storage
Pipettors should be stored in an appropriate rack or stand. This reduces the risk of
scratching or damaging the nose cone. Damage to the nose cone could result in a poor
seal to the pipet tip, which will reduce accuracy.
Testing & Calibration
You should periodically check the operation of your pipettor by checking its calibration.
Socorex air-displacement pipettors are warranted for calibration for 2 years. Under
modern quality management such as GLP / GMP, ISO-9000 or regulatory requirements,
you should test your pipettors’ calibration. If a pipettor is dropped, or you suspect any
type of damage, you should check the calibration.
Testing is generally performed gravimetrically, using an analytical balance. Calibration
is usually done at room temperature, away from drafts or direct sunlight. The actual
calibration details vary between pipettors and are listed in the instructions.