Liquid Handling
Adjustable Tip Ejection
Ejector shaft height adjustment system called Justip
™
allows a wide
selection of tips to tightly fit the nozzle.
Most Comfortable Pipetting
The Acura
®
manual
fits everybody’s hand with or without glove. Ergonomic
shape, light weight and soft plunger travel care for better comfort.
Ejector Button
Ergonomically positioned, the large surface, soft padded ejector provides for
low-pressure activation.
MinimumMaintenance / Sterilization
Easy disassembling for thorough cleaning. Sterilization achieved by auto-
claving instrument fully assembled (121°C / 250°F).
Instant Calibration
Swift-set user calibration system - with integrated key and locking
mechanism - makes tedious procedures a thing of the past. Easy and precise,
it is in full compliance with international standards and recommendations.
Safety Seal
Though easily removed when needed, the switch protection seal is
autoclaving resistant.
Cat. No.
Description
Qty / Case
W810620
Replacement seals
25
W810622
Replacement seals
100
Calibration Seal Stickers
Inaccuracy (E%)
Imprecision (CV%)
Cat. No.
Volume
Division
Min. Vol.
Mid. Vol.
Max. Vol.
Min. Vol.
Mid. Vol.
Max. Vol.
Weight (g)
Tip Style
Micropipettes Acura
®
manual 825
W810300
0.1 - 2µL
0.002µL
< ± 6.0 %
1)
< ± 4.0 % < ± 2.0 %
< 5.0 %
1)
< 3.3 % < 1.5 %
88
10µL
W810302
0.5 - 10µL
0.01µL
< ± 2.5 %
2)
< ± 1.8 % < ± 1.0 %
< 1.8 %
2)
< 1.2 % < 0.5 %
88
10µL
W810304
1 - 10YµL*
0.01µL
< ± 2.5 % < ± 1.8 % < ± 1.0 %
< 2.5 % < 1.6 % < 0.7 %
88
200µL
W810306
2 - 20µL
0.02µL
< ± 2.5 % < ± 1.8 % < ± 1.0 %
< 1.7 % < 1.1 % < 0.5 %
88
200µL
W810308
5 - 50µL
0.1µL
< ± 1.5 % < ± 1.3 % < ± 1.0 %
< 1.0 % < 0.7 % < 0.4 %
90
200µL
W810310
10 - 100µL
0.1µL
< ± 1.5 % < ± 1.2 % < ± 0.8 %
< 1.0 % < 0.6 % < 0.2 %
92
200µL
W810312
20 - 200µL
0.2µL
< ± 1.5 % < ± 1.1 % < ± 0.6 %
< 0.6 % < 0.4 % < 0.2 %
90
200µL
W810314
100 - 1000µL 1µL
< ± 1.5 % < ± 1.0 % < ± 0.5 %
< 0.5 % < 0.4 % < 0.2 %
95
1000µL
Micropipettes Acura
®
manual 826 XS
W810300-XS
0.1 - 2µL
0.002µL
< ± 6.0 %
1)
< ± 4.0 % < ± 2.0 %
< 5.0 %
1)
< 3.3 % < 1.5 %
84
10µL
W810302-XS
0.5 - 10µL
0.01µL
< ± 2.5 %
2)
< ± 1.8 % < ± 1.0 %
< 1.8 %
2)
< 1.2 % < 0.5 %
84
10µL
W810304-XS
1 - 10YµL*
0.01µL
< ± 2.5 % < ± 1.8 % < ± 1.0 %
< 2.5 % < 1.6 % < 0.7 %
85
200µL
W810306-XS
2 - 20µL
0.02µL
< ± 2.5 % < ± 1.8 % < ± 1.0 %
< 1.7 % < 1.1 % < 0.5 %
85
200µL
W810308-XS
5 - 50µL
0.1µL
< ± 1.5 % < ± 1.3 % < ± 1.0 %
< 1.0 % < 0.7 % < 0.4 %
86
200µL
W810310-XS
10 - 100µL
0.1µL
< ± 1.5 % < ± 1.2 % < ± 0.8 %
< 1.0 % < 0.6 % < 0.2 %
87
200µL
W810312-XS
20 - 200µL
0.2µL
< ± 1.5 % < ± 1.1 % < ± 0.6 %
< 0.6 % < 0.4 % < 0.2 %
86
200µL
W810314-XS
100 - 1000µL 1µL
< ± 1.5 % < ± 1.0 % < ± 0.5 %
< 0.5 % < 0.4 % < 0.2 %
88
1000µL
Performance values obtained with double-distilled water at constant temperature (± 0.5°C) comprised between 20°C and 25°C in accordance with ISO 8655.
1)
measured at 0.5µL
2)
measured at 1µL
* This pipette must be used with a 200µL pipette tip
Performance and Ordering Information - Acura
®
manual 825 and 826 XS Micropipettes